Authorised BINDER sales and service representative in Poland since 1997

Custom solutions

We adapt BINDER chambers and incubators for testing with light: we install the SUNLIGHT lamp that simulates the sun, UVA, UVB, DayLight and VIS cassettes, and LED panels. We also design for other requirements, e.g. condensation plates, to suit the specific test.

What light we install in the chamber

Each source covers a different part of the spectrum. We select them according to the standard and the purpose of the test – individually or in combination, e.g. UVA with UVB cassettes or UV cassettes with an LED panel.

The bars show the main emission range of each source according to SIMA data sheets. The intensity at the sample depends on the distance from the source and is determined for each project.
Interior of a BINDER MKF 240 chamber illuminated by a SUNLIGHT lamp mounted under the chamber ceiling
MKF 240 with SUNLIGHT lamp
BINDER MKF 240 chamber with a SUNLIGHT lamp top unit and lamp control panel

SUNLIGHT 430 W: the sun in a climate chamber

The built-in metal halide lamp reproduces the spectrum of sunlight – ultraviolet, visible light and infrared. It is used for accelerated ageing tests of materials and products, including photovoltaic cells.

radiation at a distance of 30 cm
approx. 850 W/m²
UVA
up to approx. 55 W/m²
UVB
up to approx. 10 W/m²
visible light
up to approx. 120.000 lx

1000 h ≈ 1 year

Approx. 1000 hours of exposure (approx. 42 days) corresponds to the annual dose of solar radiation under natural conditions in Poland – on average approx. 990 kWh/m² per year.

We install it in BINDER units:

  • KB 240
  • KB 720
  • KMF 240
  • KMF 720
  • MK 240
  • MK 720
  • MKF 115
  • MKF 240
  • MKF 720
  • MKFT series

Intensity vs distance

Lamp radiation with H2 filter, in W/m². We usually place samples approx. 30 cm from the lamp, no closer than 10 cm.

  1. 85030 cm
  2. 59040 cm
  3. 43550 cm
  4. 34060 cm
  5. 27570 cm
  6. 23580 cm

Filters

The filter determines the wavelength from which light reaches the sample.

  • H2 outdoor

    from 295 nm UVB, UVA, visible light and IR

  • H2 + H1 indoor

    from 315 nm UVA, visible light and IR – without UVB

  • Qz

    from 200 nm full spectrum including UVC

Maintenance

Approximate service life of the parts that are replaced during servicing.

Lamp burner
1200–⁠1500 h
Outdoor filter
15.000–⁠20.000 h
Indoor filter
4000–⁠5000 h

Fluorescent tube cassettes and LED panels

We install cassettes and panels in BINDER incubators and chambers. The details – the number of cassettes, their arrangement and the intensity at the sample – are agreed for each project.

Design
IP65
Operating temperature
0…+50 °C, with quartz covers from −5 °C*
Humidity
full range of the unit
Control
via the controller (program) or manually – depending on the series
  • Interior of a BINDER KMF 720 chamber with two UVA and UVB fluorescent tube cassettes glowing blue
    KMF 720 with UVA and UVB cassettes

    UVA and UVB cassettes

    Linear fluorescent tubes for ageing tests. On average they deliver 10–⁠20 W/m² at a distance of 10 cm from the lamps.

    UVB

    UVA

    • UVB approx. 313 nm
    • UVA approx. 365 nm
    • 10–⁠20 W/m² / 10 cm
  • DayLight cassettes

    Daylight fluorescent tubes. We install them on their own or in the same chamber as UVA and UVB cassettes.

    • daylight
    • mix with UVA and UVB
  • VIS cassettes

    Visible light for testing under constant illumination – on its own or together with a UVA cassette.

    • 400–⁠700 nm
    • 4000 K
    • 1000 lm
  • RGB LED panels

    LEDs in three colours with a narrow spectrum. The panels operate from −20 to +50 °C and also fit the KB 53 and MK 53.

    • B 465–⁠470 nm
    • G 520–⁠530 nm
    • R 625–⁠630 nm
  • DayLight LED panels

    Broad-spectrum cool white, using professional-line LEDs on a double copper substrate. From −20 to +50 °C.

    • 6000–⁠6500 K
    • 1080–⁠1200 lm/m
    • up to 50.000 h

The cassettes fit: KB, KMF and KBF 115, 240 and 720, KBW, KBWF, MK 115, MKT, MKF 115, 240 and 720. LED panels also fit KB 53 and MK 53.
* At sub-zero temperatures, the radiation intensity of the cassettes drops by approx. 50 %.

SIMA installations

BINDER units in which we have installed lamps, cassettes and LED lighting. Click a photo to enlarge it.

BINDER lighting solutions

Some requirements are met by the manufacturer’s standard modules, while BINDER builds non-standard versions to order. We will advise you on which solution suits your study.

  • BINDER KBF PRO 720 plant growth chamber with LED module and plants

    LED module for plant illumination

    • 16 LED strips
    • 3 colours
    • 10–⁠100 %

    Turns KBF and KBF PRO chambers and KB PRO incubators (sizes 260, 470, 720, 1060 and 1600) into plant growth chambers. The 16 LED strips emit light in three colours: cool white, warm white and deep red; each colour can be dimmed separately in the range 10–⁠100 %.

  • BINDER KBF PRO 720 chamber for photostability testing with ICH light module

    ICH light modules

    • ICH Q1B, option 2
    • VIS + UVA
    • ICH-LQC

    For photostability testing according to ICH Q1B (option 2) in KBF and KBF PRO 260, 470 and 720 chambers and, without humidity control, in KB PRO 260, 470 and 720. They have separate cassettes: VIS with LEDs and UVA with fluorescent tubes. The ICH-LQC version controls the light dose.

  • LED strip kits

    • 30, 50 or 90 cm strips
    • +2 strips

    For illuminating the interior of BINDER climate chambers with low-intensity light. The basic kit (30, 50 or 90 cm strips with a control unit) can be extended with two strips.

  • BINDER made-to-order versions

    • from 1 unit
    • Tuttlingen

    The BINDER Customized Solutions department builds special versions based on series-production units, even as single units, at the Tuttlingen plant. For lighting, these include, for example, cassettes with black light and daylight, cassettes with light sensors on each level, or LED strips.

What to prepare when describing your project

This information will help us assess which solution is possible.

  1. Purpose of the study

    What you want to test, e.g. a material’s resistance to UV radiation or plant growth under specific light.

  2. Standards and procedures

    The number of the standard, guideline or internal procedure you work to.

  3. Parameters

    Temperature and humidity range, type of radiation or light (waveband, intensity), exposure time and cycles.

  4. Samples

    Type, number, dimensions and weight, and how they are arranged in the chamber.

  5. Equipment

    Model and serial number if you already have a BINDER unit; if you are looking for a new one – the installation site and available power supply.

  6. Documentation and deadline

    Whether you need measurements (e.g. temperature or light distribution), IQ/OQ qualification or data logging, and by when.

Is a standard model not enough?

Describe your study – purpose, standard and parameters. We will check whether a BINDER module is sufficient or a made-to-order version is needed.

Contact

Sales office and service

SIMA
ul. Pokrzywnicka 3a
04-320 Warszawa

Show on map

Opening hours
24h/7

Write to us

Choose a topic and describe what you need. We will reply by email or phone.

What is your message about?

The data controller is „SIMA” s.c. P. Sikora, M. Gniadzik, ul. Pokrzywnicka 3a, 04-320 Warszawa. We will use the data from the form to reply to your message. More in our privacy policy.