Custom solutions
We adapt BINDER chambers and incubators for testing with light: we install the SUNLIGHT lamp that simulates the sun, UVA, UVB, DayLight and VIS cassettes, and LED panels. We also design for other requirements, e.g. condensation plates, to suit the specific test.
What we install in BINDER chambers
approx. 850 W/m²SUNLIGHT 430 Wat a distance of 30 cm, for ageing tests10–20 W/m²UVA, UVB, DayLight, VIS cassettesUV at a distance of 10 cm from the lampsR · G · BRGB and DayLight LED panelsalso white 6000–6500 K, from −20 to +50 °C authorised representative since 1997
What light we install in the chamber
Each source covers a different part of the spectrum. We select them according to the standard and the purpose of the test – individually or in combination, e.g. UVA with UVB cassettes or UV cassettes with an LED panel.
- SUNLIGHT 430 Wmetal halide lamp, with H2 filter from 295 nm
- UVB cassettesfluorescent tubes, peak approx. 313 nm
- UVA cassettesfluorescent tubes, peak approx. 365 nm
- DayLight cassettesdaylight fluorescent tubes
- VIS cassettes400–700 nm, colour temperature 4000 K
- RGB LED panelsblue, green and red
- DayLight LED panelscool white 6000–6500 K


SUNLIGHT 430 W: the sun in a climate chamber
The built-in metal halide lamp reproduces the spectrum of sunlight – ultraviolet, visible light and infrared. It is used for accelerated ageing tests of materials and products, including photovoltaic cells.
- radiation at a distance of 30 cm
- approx. 850 W/m²
- UVA
- up to approx. 55 W/m²
- UVB
- up to approx. 10 W/m²
- visible light
- up to approx. 120.000 lx
1000 h ≈ 1 year
Approx. 1000 hours of exposure (approx. 42 days) corresponds to the annual dose of solar radiation under natural conditions in Poland – on average approx. 990 kWh/m² per year.
We install it in BINDER units:
- KB 240
- KB 720
- KMF 240
- KMF 720
- MK 240
- MK 720
- MKF 115
- MKF 240
- MKF 720
- MKFT series
Intensity vs distance
Lamp radiation with H2 filter, in W/m². We usually place samples approx. 30 cm from the lamp, no closer than 10 cm.
- 85030 cm
- 59040 cm
- 43550 cm
- 34060 cm
- 27570 cm
- 23580 cm
Filters
The filter determines the wavelength from which light reaches the sample.
H2 outdoor
from 295 nm UVB, UVA, visible light and IR
H2 + H1 indoor
from 315 nm UVA, visible light and IR – without UVB
Qz
from 200 nm full spectrum including UVC
Maintenance
Approximate service life of the parts that are replaced during servicing.
- Lamp burner
- 1200–1500 h
- Outdoor filter
- 15.000–20.000 h
- Indoor filter
- 4000–5000 h
Fluorescent tube cassettes and LED panels
We install cassettes and panels in BINDER incubators and chambers. The details – the number of cassettes, their arrangement and the intensity at the sample – are agreed for each project.
- Design
- IP65
- Operating temperature
- 0…+50 °C, with quartz covers from −5 °C*
- Humidity
- full range of the unit
- Control
- via the controller (program) or manually – depending on the series

KMF 720 with UVA and UVB cassettes UVA and UVB cassettes
Linear fluorescent tubes for ageing tests. On average they deliver 10–20 W/m² at a distance of 10 cm from the lamps.
UVB
UVA
- UVB approx. 313 nm
- UVA approx. 365 nm
- 10–20 W/m² / 10 cm
DayLight cassettes
Daylight fluorescent tubes. We install them on their own or in the same chamber as UVA and UVB cassettes.
- daylight
- mix with UVA and UVB
VIS cassettes
Visible light for testing under constant illumination – on its own or together with a UVA cassette.
- 400–700 nm
- 4000 K
- 1000 lm
RGB LED panels
LEDs in three colours with a narrow spectrum. The panels operate from −20 to +50 °C and also fit the KB 53 and MK 53.
- B 465–470 nm
- G 520–530 nm
- R 625–630 nm
DayLight LED panels
Broad-spectrum cool white, using professional-line LEDs on a double copper substrate. From −20 to +50 °C.
- 6000–6500 K
- 1080–1200 lm/m
- up to 50.000 h
The cassettes fit: KB, KMF and KBF 115, 240 and 720, KBW, KBWF, MK 115, MKT, MKF 115, 240 and 720. LED panels also fit KB 53 and MK 53.
* At sub-zero temperatures, the radiation intensity of the cassettes drops by approx. 50 %.
SIMA installations
BINDER units in which we have installed lamps, cassettes and LED lighting. Click a photo to enlarge it.
MKF 240 with SUNLIGHT lamp
KB/KMF 240 with SUNLIGHT lamp and filter
KB 240 with SUNLIGHT 430 W lamp
KMF 115 with 4 × 15 W UVA cassette
KB 240/KMF with SUNLIGHT lamp – indoor
MKF with SUNLIGHT lamp switched on
BINDER incubators with additional LED lightingNot just light
We design condensation plates and other components to meet the requirements of a specific study.
Describe your requirements
BINDER lighting solutions
Some requirements are met by the manufacturer’s standard modules, while BINDER builds non-standard versions to order. We will advise you on which solution suits your study.

LED module for plant illumination
- 16 LED strips
- 3 colours
- 10–100 %
Turns KBF and KBF PRO chambers and KB PRO incubators (sizes 260, 470, 720, 1060 and 1600) into plant growth chambers. The 16 LED strips emit light in three colours: cool white, warm white and deep red; each colour can be dimmed separately in the range 10–100 %.

ICH light modules
- ICH Q1B, option 2
- VIS + UVA
- ICH-LQC
For photostability testing according to ICH Q1B (option 2) in KBF and KBF PRO 260, 470 and 720 chambers and, without humidity control, in KB PRO 260, 470 and 720. They have separate cassettes: VIS with LEDs and UVA with fluorescent tubes. The ICH-LQC version controls the light dose.
LED strip kits
- 30, 50 or 90 cm strips
- +2 strips
For illuminating the interior of BINDER climate chambers with low-intensity light. The basic kit (30, 50 or 90 cm strips with a control unit) can be extended with two strips.
BINDER made-to-order versions
- from 1 unit
- Tuttlingen
The BINDER Customized Solutions department builds special versions based on series-production units, even as single units, at the Tuttlingen plant. For lighting, these include, for example, cassettes with black light and daylight, cassettes with light sensors on each level, or LED strips.
What to prepare when describing your project
This information will help us assess which solution is possible.
Purpose of the study
What you want to test, e.g. a material’s resistance to UV radiation or plant growth under specific light.
Standards and procedures
The number of the standard, guideline or internal procedure you work to.
Parameters
Temperature and humidity range, type of radiation or light (waveband, intensity), exposure time and cycles.
Samples
Type, number, dimensions and weight, and how they are arranged in the chamber.
Equipment
Model and serial number if you already have a BINDER unit; if you are looking for a new one – the installation site and available power supply.
Documentation and deadline
Whether you need measurements (e.g. temperature or light distribution), IQ/OQ qualification or data logging, and by when.
Is a standard model not enough?
Describe your study – purpose, standard and parameters. We will check whether a BINDER module is sufficient or a made-to-order version is needed.