BINDER stability test chambers
BINDER climate chambers for stability testing maintain constant temperature and humidity during long-term continuous operation, uniformly throughout the interior. They are used for testing pharmaceuticals according to ICH Q1A and Q1B and for shelf-life testing of cosmetics, food and materials, while large batches of samples will fit in a walk-in chamber. We will help you choose a model and, after purchase, take care of service, calibration, temperature and humidity distribution measurements according to DIN 12880 and IQ/OQ qualification.
Operating temperature range−20 to +100 °C

Applications
- Drug stability studies according to ICH Q1A(R2): long-term, accelerated and stress testing
- Photostability testing according to ICH Q1B
- Shelf-life testing of cosmetics and food
- Material testing and THB 85/85 testing (KBF PRO)
- Accelerated ageing of medical device packaging (ASTM F1980)
- Storage of large sample batches under constant climatic conditions

KBF · KBF PRO · KBF-S ECO
Climate chambers for stability testing
For drug stability studies according to ICH Q1A: long-term, accelerated and stress studies, as well as for shelf-life testing of cosmetics and food. The chambers maintain the set temperature and humidity uniformly throughout the interior, even when fully loaded. There are three series to choose from, with volumes from 127 to 1.610 l.
- APT.line™ preheating chamber: uniform climate even when fully loaded
- Steam humidification and capacitive humidity sensor (KBF and KBF PRO)
- Independent class 3.3 temperature safety device (DIN 12880) with alarm (KBF and KBF PRO)
- Data logger with readout via USB, Ethernet interface; optional APT-COM™ GLP Edition (GLP, FDA 21 CFR Part 11)
- Cooling with R-600a (KBF) or R-290 (KBF PRO) refrigerant; KBF-S ECO cools with a Peltier module, without refrigerant
- Humidification water: demineralised, from a fixed water supply or from a canister (KBF-S ECO: canister only)
- Temperature
- KBF: 0 to +70 °CKBF PRO: −20 to +100 °CKBF-S ECO: 0 to +70 °C (max. 24–28 °C below ambient temperature)
- Humidity
- KBF and KBF-S ECO: 10–80 %KBF PRO: 10–98 %
- Humidity control in the temperature range
- KBF: +10 to +70 °CKBF PRO: +10 to +90 °CKBF-S ECO: +5 to +70 °C
- Constancy over time (25 °C and 60 % RH, 40 °C and 75 % RH)
- temperature: ±0.1 Khumidity: KBF ±1.5 % RH, KBF PRO ±1 % RH, KBF-S ECO ±0.3–0.4 % RH
- Models
- KBF and KBF PRO: 130, 260, 470, 720, 1060, 1600KBF-S ECO: 240, 400, 720, 1020
KBF Chambers with a wide temperature and humidity range
From 0 to +70 °C and from 10 to 80 % RH: for long-term, accelerated and stress testing according to ICH. KBF 130 units can be stacked; larger models stand on castors with brakes.
| Model | Capacity | Temperature | Humidity | Energy consumption(25 °C, 60 % RH) | External dimensions(W × H × D, mm) | Data sheet | Enquiry |
|---|---|---|---|---|---|---|---|
| KBF 130 | 127 l | from 0 to +70 °C | 10–80 % | 215 Wh/h | 900 × 1005 × 710 | PDF, 0.7 MB | Enquire |
| KBF 260 | 262 l | from 0 to +70 °C | 10–80 % | 215 Wh/h | 930 × 1396 × 900 | PDF, 0.8 MB | Enquire |
| KBF 470 | 472 l | from 0 to +70 °C | 10–80 % | 265 Wh/h | 930 × 1951 × 900 | PDF, 0.7 MB | Enquire |
| KBF 720 | 727 l | from 0 to +70 °C | 10–80 % | 360 Wh/h | 1285 × 1951 × 900 | PDF, 0.7 MB | Enquire |
| KBF 1060 | 1070 l | from 0 to +70 °C | 10–80 % | 365 Wh/h | 1285 × 1951 × 1175 | PDF, 0.8 MB | Enquire |
| KBF 1600 | 1610 l | from 0 to +70 °C | 10–80 % | 395 Wh/h | 1795 × 1951 × 1175 | PDF, 0.8 MB | Enquire |
KBF PRO Chambers with an extended temperature and humidity range
From −20 to +100 °C and from 10 to 98 % RH, so also for material testing and THB testing at +85 °C and 85 % RH.
| Model | Capacity | Temperature | Humidity | Energy consumption(25 °C, 60 % RH) | External dimensions(W × H × D, mm) | Data sheet | Enquiry |
|---|---|---|---|---|---|---|---|
| KBF PRO 130 | 127 l | from −20 to +100 °C | 10–98 % | 235 Wh/h | 900 × 1005 × 710 | PDF, 0.7 MB | Enquire |
| KBF PRO 260 | 262 l | from −20 to +100 °C | 10–98 % | 235 Wh/h | 930 × 1396 × 900 | PDF, 0.8 MB | Enquire |
| KBF PRO 470 | 472 l | from −20 to +100 °C | 10–98 % | 245 Wh/h | 930 × 1951 × 900 | PDF, 0.7 MB | Enquire |
| KBF PRO 720 | 727 l | from −20 to +100 °C | 10–98 % | 390 Wh/h | 1285 × 1951 × 900 | PDF, 0.8 MB | Enquire |
| KBF PRO 1060 | 1070 l | from −20 to +100 °C | 10–98 % | 400 Wh/h | 1285 × 1951 × 1175 | PDF, 0.8 MB | Enquire |
| KBF PRO 1600 | 1610 l | from −20 to +100 °C | 10–98 % | 490 Wh/h | 1795 × 1951 × 1175 | PDF, 0.8 MB | Enquire |
KBF-S ECO Chambers with thermoelectric cooling (Peltier module)
It cools with a Peltier module, without refrigerant, and consumes 55–90 Wh/h at 25 °C and 60 % RH. It cools down to 24–28 °C below ambient temperature and draws humidification water from a canister.
| Model | Capacity | Temperature | Humidity | Energy consumption(25 °C, 60 % RH) | External dimensions(W × H × D, mm) | Data sheet | Enquiry |
|---|---|---|---|---|---|---|---|
| KBF-S ECO 240 | 247 l | from 0 to +70 °C (max. 28 °C below ambient) | 10–80 % | 55 Wh/h | 932 × 1461 × 796 | PDF, 0.6 MB | Enquire |
| KBF-S ECO 400 | 400 l | from 0 to +70 °C (max. 26 °C below ambient) | 10–80 % | 60 Wh/h | 925 × 1946 × 796 | PDF, 0.5 MB | Enquire |
| KBF-S ECO 720 | 700 l | 0 to +70 °C (max. 24 °C below ambient) | 10–80 % | 90 Wh/h | 1254 × 1925 × 885 | PDF, 0.7 MB | Enquire |
| KBF-S ECO 1020 | 1020 l | 0 to +70 °C (max. 24 °C below ambient) | 10–80 % | 90 Wh/h | 1254 × 1925 × 1146 | PDF, 0.7 MB | Enquire |
External dimensions as specified by BINDER: housing without accessories or connections.
BINDER materials to download
See also: Plant growth chambers: KBF and KBF PRO with LED module for plant illumination

KBF · KBF PRO · KB PRO with ICH module
Photostability test chambers
The ICH light module turns a KBF or KBF PRO chamber (sizes 260, 470 and 720) into a photostability test chamber according to ICH Q1B (option 2), and for tests without humidity control it also fits the KB PRO cooled incubator. Samples are exposed in sequence to visible light from LEDs and to UVA radiation from fluorescent tubes. A complete ICH test can be carried out in two days.
- Separate VIS and UVA cassettes; after exposure to one source, the samples are moved under the other
- LQC version: VIS and UVA sensors monitor the dose and switch off the light source once it has been reached
- Light intensity limited to avoid thermal effects during the test
- Lighting controlled by a time programme from the chamber controller
- The module can be retrofitted later, on site; it fits current-generation chambers (E7)
- Optional APT-COM™ GLP Edition: parameter recording compliant with GLP and FDA 21 CFR Part 11
- Compliance
- ICH Q1B, option 2
- Light sources
- VIS: LEDs, spectrum close to D65UVA: fluorescent tubes
- Doses required by ICH Q1B
- visible light: at least 1.2 million lx·hUVA (320–400 nm): at least 200 Wh/m²
- Modules
- ICH 260/470 and ICH 720with dose control: ICH-LQC 260/470 and ICH-LQC 720
| Base unit | Light module | Temperature | Humidity |
|---|---|---|---|
| KBF 260, KBF 470 | ICH 260/470 or ICH-LQC 260/470 | from +20 to +50 °C | 20–50 % |
| KBF 720 | ICH 720 or ICH-LQC 720 | from +20 to +50 °C | 20–50 % |
| KBF PRO 260, KBF PRO 470 | ICH 260/470 or ICH-LQC 260/470 | from +10 to +50 °C | 10–80 % |
| KBF PRO 720 | ICH 720 or ICH-LQC 720 | from +10 to +50 °C | 10–80 % |
| KB PRO 260, KB PRO 470 (cooled incubator) | ICH 260/470 or ICH-LQC 260/470 | from +5 to +60 °C | not controlled |
| KB PRO 720 (cooled incubator) | ICH 720 or ICH-LQC 720 | from +5 to +60 °C | not controlled |
BINDER materials to download

WIC 1 · WIC 2 · WIC 3
Walk-in chambers
Chambers you can walk into, for long-term and accelerated stability studies according to ICH Q1A(R2) and for storing large batches of samples. Three standard sizes with volumes of approx. 12.5 m³, 18.5 m³ and 24.5 m³, and up to 60 m³ in the modular version, with dimensions, doors and windows adapted to the room.
- Stainless steel interior and floor, floor load capacity 50.000 N/m²
- Walls made of panels with 100 mm PU foam (U = 0.21 W/m²K)
- LED lighting with motion sensor
- Air-conditioning unit at the rear of the chamber (optionally at the front), with service access from outside
- Refrigerant with a GWP of 547
- Additional equipment: stainless steel shelving, ramp, 30 and 50 mm access ports, PVC strip curtain, APT-COM™-compatible monitoring system
- Temperature
- +5 to +50 °C, optionally from −20 °C
- Humidity
- 20–90 %
- Capacity
- WIC 1: 12.5 m³, WIC 2: 18.5 m³, WIC 3: 24.5 m³modular version: up to 60 m³
- Uniformity and constancy
- temperature: ±1.5 K spatially, ±0.5 K over timehumidity: ±2.5 % RH over time
- Power supply
- 230 V, 50 Hz, 1 phase; power 2.7 kW

WIC Walk-in chambers in standard sizes
Externally, each is 2.6 m wide and 2.47 m high, with a length of 2.5, 3.6 or 4.7 m. The interior is 2.27 m high.
| Model | Capacity | Temperature | Humidity | External dimensions(W × H × D, mm) | Data sheet | Enquiry |
|---|---|---|---|---|---|---|
| WIC 1 | 12.530 l | from +5 to +50 °C | 20–90 % | 2600 × 2470 × 2500 | PDF, 1.3 MB | Enquire |
| WIC 2 | 18.523 l | from +5 to +50 °C | 20–90 % | 2600 × 2470 × 3600 | PDF, 1.9 MB | Enquire |
| WIC 3 | 24.516 l | from +5 to +50 °C | 20–90 % | 2600 × 2470 × 4700 | PDF, 1.7 MB | Enquire |
External dimensions as specified by BINDER: housing without accessories or connections.
Service, calibration and qualification
We are an authorised BINDER service provider. We look after equipment both under warranty and after the warranty has expired, calibrate it, measure temperature distribution according to DIN 12880 and carry out IQ/OQ qualification.
Contact
Sales office and service
SIMA
ul. Pokrzywnicka 3a
04-320 Warszawa
- Opening hours
- 24h/7
Write to us
Choose a topic and describe what you need. We will reply by email or phone.
The data controller is „SIMA” s.c. P. Sikora, M. Gniadzik, ul. Pokrzywnicka 3a, 04-320 Warszawa. We will use the data from the form to reply to your message. More in our privacy policy.